For full functionality of this site it is necessary to enable JavaScript. Here are the instructions how to enable JavaScript in your web browser.

Scientific Program

FP447

1939

Dr.Vineeth B Bhimshetty
Membership No Author Name Email Mobile
3600 Dr.Mahitha C S mahithasathyananda@gmail.com 8277412696
Cataract

Comparison of ARGOS and OA-2000 Optical Biometers for IOL Calculation

Purpose: To compare the agreement and clinical interchangeability of ocular biometric measurements obtained with the ARGOS (Alcon) and OA-2000 (Tomey) optical biometers for intraocular lens (IOL) power calculation.
Methods: In this prospective study, 200 eyes underwent biometry with both devices. Parameters assessed included K1, K2, axial length (AL), anterior chamber depth (ACD), lens thickness (LT), and white-to-white (WTW). Agreement was evaluated using Bland–Altman analysis, Pearson correlation, and paired t-test.
Results: Mean AL was 23.14 mm (ARGOS) and 23.10 mm (OA-2000), with a mean offset of +0.05 mm. ACD and LT offsets were +0.07 mm and −0.14 mm. AL and keratometry showed excellent correlation (AL r=0.99, Km r=0.98) with narrow AL agreement limits (−0.24 to +0.33 mm).
Conclusion: ARGOS and OA-2000 showed excellent agreement for AL and keratometry, supporting their interchangeable use for IOL power calculation and cataract surgery planning.


Membership No Author Name Email Mobile
2108 Dr.Vijayalakshmi vijayalakshmiptl@gmail.com 9620676762
1939 Dr.Vineeth B Bhimshetty vins.vineeth@gmail.com 9620676761
1877 Dr.Vinut Kumar V Anandi vinutanandi@gmail.com 09986152298